PPP-NCST datasheet
AFM Probe Type:
PPP-NCST
Product Description:
The PointProbe® Plus (PPP) combines the well-known features of the proven PointProbe® series such as high application versatility and compatibility with most commercial SPMs with a further reduced and more reproducible tip radius as well as a more defined tip shape. The excellent tip radius and the minimized variation in tip shape provide more reproducible images and enhanced resolution.
NANOSENSORS™ PPP-NCST AFM probes are designed for non-contact or soft tapping mode imaging. The combination of soft cantilever and fairly high resonance frequency enables stable and fast measurements with reduced tip-sample interaction. This feature significantly reduces tip wear and sample wear at the same time
The probe offers unique features:
- excellent tip radius of curvature
- highly doped to dissipate static charge
- chemically inert
- high mechanical Q-factor for high sensitivity
Application Modes:
AFM Probe Specifications:
AFM Cantilevers:
| Cant. | Length | Width | Thickness | Force Const. | Res. Freq. |
|---|---|---|---|---|---|
| 1 | 150 µm(140 - 160 µm)* | 27 µm(19.5 - 34.5 µm)* | 2.8 µm(1.8 - 3.8 µm)* | 7.4 N/m(1.2 - 29 N/m)* | 160 kHz(75 - 265 kHz)* |
AFM Tips:
| on Cant. | Shape | Tip Radius |
|---|---|---|
| 1 | Standard | <7 nm( < 10 nm guaranteed ) |
* guaranteed range
Coating:
None
Purchasing Info:
Product Availability: |
From stock |
Price Description: |
Does not include VAT or customs duties. |